2. MSS-5



FEATURES

Calibrated for the specific anaesthetic drug.
Calibrated 0.25 to 5% (Sevoflurane 8%) using a "Laser Refractometer"
Temperature, flow and duration of use compensation.
Safety lock to help prevent accidental turn-on of dial.

SPECIFICATION
Calibrated at 21 deg C and 35 deg C +/- 1 deg C.
Calibration range: 0.25% - 0.5% through 5% (Sevoflurane 8%)v.v.
Fluctuations in temperature, flow rate and duration of use is automatically compensated for during use.
When used with I.P.P.V the variation in output is negligible.
Nominal Working Range : Fresh gas flow - 200 cc to 15 litre/min.
Ambient temperature - 15 deg C to 40 deg C.

Resistance to Gas Flow:
5 cm wg. at "OFF" setting at 5 litre/min oxygen.
21-29 cm wg. when delivering vapour in 5 litre/min oxygen at 21oC.
Liquid Capacity:
Amount of anaesthetic drug to fully charge the vaporiser - 200ml
Amount retained by the wick system - 55ml.
Weight & Dimensions:
(Screw fill)
Weight 7.0 kg.
Height 200mm
Depth 200mm
Width 120mm

Service Options
Option 1. Planned Preventative Maintenance - Once every 3 years the vaporiser should be returned for service incl calibration check
Option 2. After the first 3 years from new, check calibration. Then check the calibration once every following 6 months. If the calibration shows evidence of deterioration, the vaporiser should be returned for service. This requires records to be kept for tracking.
PLEASE NOTE:- HALOTHANE Vaporisers should be serviced annually.

ORDERING INFORMATION

When ordering please specify which variant is required. Drug type: Isoflurane 5%, Halothane 5%, Sevoflurane 8%, Mounting system is Seletatec® , with Interlock. Filling Type: Keyed filling, Screw filling. The filling adaptor is NOT included but can be supplied on request* FOR FURTHER INFORMATION ON PRICE, AVAILABILITY, AFTER SALES SERVICE AND OTHER PRODUCTS PLEASE CONTACT: MSS Int'l Ltd.
 

Website developed by Cultrix Ltd.